Textbook / Chapter 15 / Clinical examination, investigations and imaging

Ultrasound, biometry and corneal topography

This lesson covers the tests that measure the eye: ultrasound for seeing behind opaque media and for measuring length, optical biometry, and the instruments that measure the cornea's curvature, shape, thickness and endothelium. They decide the power of a lens implant, detect keratoconus and warn of a cornea that may not survive surgery.

10 minute read6 linked questions

Ultrasound: A-scan and B-scan

Ultrasound is sound above the range of hearing. A crystal in the probe sends short pulses into the eye and listens for echoes. An echo returns from every boundary between tissues that carry sound differently, and the delay before it arrives gives the depth of that boundary. Higher frequencies give finer detail but penetrate less. The eye is small and superficial, so ophthalmic probes run at about 10 MHz, well above the frequencies used for the abdomen.

The echoes can be displayed in two ways. An A-scan (amplitude) sends a single thin beam and plots the echoes as spikes along a line. Spike height is echo strength and distance along the line is depth. It is a one-dimensional measurement, used for distances. A B-scan (brightness) sweeps the beam through an arc and plots each echo as a dot whose brightness is its strength, building a two-dimensional slice of the eye and orbit. It is used to look at shape and position.

An A-scan along the axis of a normal eye shows a tall spike from the cornea, two from the front and back of the lens, a flat silent stretch through the vitreous, and a tall spike from the retina followed by descending echoes from the sclera and orbital fat. Ultrasound biomicroscopy uses much higher frequencies, 35 to 50 MHz, to image only the front few millimetres in fine detail: the angle, the ciliary body and what lies behind the iris.

A-scanB-scan
DisplaySpikes along one lineTwo-dimensional grey image
Tells youDistances, and the reflectivity inside a lesionShape, position and movement
Main usesAxial length for lens calculation; characterising tumoursExamining the posterior segment when the view is blocked
A-scan and B-scan compared

What B-scan is used for

The main indication is opaque media. When a dense cataract, a corneal scar or a vitreous haemorrhage hides the fundus, B-scan shows whether the retina is attached. The probe is placed on the closed lid with gel, and the patient moves the eye so the examiner can watch how any membrane moves.

The classic problem is a membrane seen in the vitreous after a haemorrhage. It may be a detached retina, which needs surgery, or the detached posterior face of the vitreous, which usually does not.

Retinal detachmentPosterior vitreous detachment
MembraneThick, bright and foldedThin and faint
Attachment at the optic discRemains attachedFree of the disc when complete
Movement after the eye movesStiff, limited after-movementLoose, wafting after-movement
A-scan spikeAs tall as the scleral spikeLower
When the gain is turned downStill visibleDisappears
Retinal detachment and posterior vitreous detachment on ultrasound

B-scan also characterises masses. A choroidal melanoma is a solid dome, or a mushroom (collar-stud) shape once it has broken through Bruch membrane. Its tightly packed cells give low to medium internal reflectivity, so the inside looks dark and hollow, and it may scoop out the underlying choroid. A choroidal haemangioma is a dome with high internal reflectivity. Ultrasound measures tumour height and base, which guide treatment and follow-up.

  • Posterior scleritis: thickened coats with fluid behind the globe that joins the dark shadow of the optic nerve to form the T-sign.
  • Optic disc drusen: bright, calcified spots in the nerve head that cast a shadow. This separates a harmless lumpy disc from true disc swelling.
  • Intraocular foreign body: a very bright echo with shadowing behind it.
  • Retinoblastoma: a mass containing bright flecks of calcium.

Biometry: measuring axial length

Biometry is the measurement of the eye before cataract surgery, chiefly its axial length and corneal power, from which the power of the lens implant is calculated. The formulae and the NICE recommendations are taught in the cataract chapter. This section covers how the measurements are made.

Ultrasound A-scan biometry times the echo from the retina and converts time to distance using the speed of sound. That speed differs between tissues: about 1532 metres per second in aqueous and vitreous and about 1641 in the cornea and lens. The machine must be told what the eye contains. An eye filled with silicone oil after retinal surgery carries sound far more slowly, and unless this is corrected the eye measures falsely long.

In the contact (applanation) method the probe touches the cornea. Any pressure indents the cornea and shortens the reading, and a short reading leads to an implant that is too strong. In the immersion method the probe sits in a small bath of saline above the eye and never touches it, which removes the compression error. Ultrasound measures to the inner surface of the retina along the line the probe happens to point, which in a highly myopic eye with a bulging posterior pole may miss the fovea.

Optical biometry has replaced ultrasound as the first choice. The original method, partial coherence interferometry, uses infrared light of 780 nm and the same interference principle as OCT. The patient looks at a fixation light, so the measurement runs along the visual axis to the fovea, and nothing touches the eye. The same machine measures corneal curvature and anterior chamber depth. Its weakness is that light must get through: very dense cataracts and dense posterior subcapsular opacities can defeat it, and then ultrasound is used.

Keratometry, topography and tomography

The front of the cornea, with its tear film, is the most powerful refracting surface of the eye, and it behaves as a convex mirror. A steeper mirror gives a smaller reflection. Every instrument that measures corneal curvature starts from that fact.

A keratometer projects a pattern onto the cornea and measures the size of its reflection in two meridians at right angles, the steepest and the flattest. It reports them as K readings in dioptres. Normal central values lie between about 40 and 47 D, with a mean of 43 D. The difference between the two is the corneal astigmatism. Keratometry samples only a small ring about 3 mm across at the centre and assumes the cornea between is regular. That is enough for a healthy cornea and a lens calculation. It says nothing about the periphery and is unreliable on an irregular surface.

Corneal topography extends the idea across the whole surface. A Placido disc, a set of concentric bright rings, is reflected from the cornea and photographed. Where the cornea is steep the rings are crowded together, and where it is flat they spread apart. Software converts the ring spacing to curvature at thousands of points and draws a colour map. By convention warm colours (red, orange) mark steep, powerful areas and cool colours (blue) mark flat ones. Regular astigmatism appears as a symmetrical bow tie along the steep meridian. Because the image is a reflection from the tear film, a dry or irregular surface degrades it, and only the front surface is measured.

Corneal tomography images the cornea in cross-section instead of using a reflection. A rotating Scheimpflug camera, a scanning slit or anterior segment OCT takes a series of optical slices and reconstructs both the front and the back surface. From these it derives elevation maps of each surface and a map of thickness at every point. Topography gives the shape of the front surface. Tomography gives the whole cornea in three dimensions.

KeratometryPlacido topographyTomography
Area measuredCentral ring about 3 mm acrossMost of the front surfaceFront and back surfaces
MethodReflection, two meridiansReflection of ringsOptical cross-sections
Posterior surfaceNoNoYes
Thickness mapNoNoYes
Main useLens calculation, contact lens fittingIrregular astigmatism, screeningKeratoconus, screening before refractive surgery
Measuring the shape of the cornea

Keratoconus on the map

In keratoconus the cornea thins and bulges forward into a cone, usually below the centre. Early disease looks normal at the slit lamp, so mapping makes the diagnosis. The earliest clue in clinic is a scissoring reflex on retinoscopy and irregular astigmatism that keeps changing.

  • Inferior steepening: a hot spot below the centre of the map.
  • An asymmetric bow tie whose two lobes are unequal and not in line with each other.
  • High K readings: central power above about 47 D is one of the classic screening criteria.
  • Inferior-superior asymmetry: the lower cornea more than about 1.4 D steeper than the upper.
  • On tomography, forward bulging of the posterior surface, often an early change, and a thinnest point that is displaced downwards and coincides with the steepest point.

Serial maps show whether the disease is progressing, which is the trigger for corneal collagen cross-linking. The same maps screen candidates for laser refractive surgery, because removing tissue from an unrecognised early cone can cause the cornea to bulge afterwards (ectasia).

Pachymetry and specular microscopy

Pachymetry measures corneal thickness. An ultrasound pachymeter is a small probe touched to the anaesthetised cornea that times the echo from its back surface. Optical methods (tomography and OCT) do the same without contact and give a map. The central cornea is a little over half a millimetre thick. Three uses matter.

  • Glaucoma: Goldmann tonometry reads high on a thick cornea and low on a thin one, so central corneal thickness is measured when glaucoma is diagnosed.
  • Refractive surgery and keratoconus: enough tissue must remain after laser, and a thin or thinning cornea suggests ectasia.
  • Endothelial function: a failing endothelium lets the stroma swell, so a thickening cornea is a measure of decompensation.

The corneal endothelium is a single layer of hexagonal cells that pumps fluid out of the stroma and keeps it clear. The cells do not divide in any useful number in adult life. When cells are lost, their neighbours enlarge and slide to fill the gap, so the count per unit area falls and the regular pattern is lost.

Specular microscopy photographs this layer using the mirror-like reflection from the interface between endothelium and aqueous. Software counts the cells and reports three numbers. Endothelial cell density in a healthy adult is about 2500 to 3000 cells per square millimetre and declines slowly with age. Polymegathism is variation in cell size, reported as a coefficient of variation. Pleomorphism is loss of the six-sided shape: a healthy cornea is expected to have at least 60% hexagonal cells. In Fuchs endothelial dystrophy dark gaps called guttae appear in the mosaic, beginning centrally. A low count before cataract surgery warns that the cornea may decompensate afterwards, and shapes the consent discussion and surgical technique.

3D atlasThe endothelium, the single cell layer that specular microscopy counts.

Summary

  • A-scan ultrasound plots echoes as spikes along one line and measures distance; B-scan builds a two-dimensional image.
  • B-scan is used when the fundus cannot be seen, above all to find a retinal detachment behind a vitreous haemorrhage or dense cataract.
  • A detached retina is a thick, stiff membrane that stays attached at the disc; a detached vitreous face is thin, mobile and free of the disc.
  • Choroidal melanoma is a dome or mushroom with low to medium internal reflectivity; posterior scleritis gives the T-sign.
  • Contact A-scan biometry can compress the cornea and under-measure axial length; immersion ultrasound and optical biometry avoid this.
  • Optical biometry measures along the visual axis without contact but can fail through very dense cataract.
  • Keratometry measures only a small central ring in two meridians; topography maps the front surface and tomography maps both surfaces and thickness.
  • Keratoconus shows inferior steepening, an asymmetric bow tie, high K readings, posterior surface bulging and a displaced thinnest point.
  • Central corneal thickness is needed to interpret Goldmann tonometry.
  • Specular microscopy counts endothelial cells: a healthy adult has about 2500 to 3000 per square millimetre with at least 60% hexagonal.

Check yourself

Quick recall on this lesson. These are easier than the question bank.

1. A vitreous haemorrhage hides the fundus. Which test shows whether the retina is detached?
2. What is the usual error of contact A-scan biometry?
3. Which finding on corneal topography suggests keratoconus?
4. What does specular microscopy examine?
Practise 6 questions on this lesson

Written for revision. It is not a source of clinical advice, and guidance changes.