Optical instruments, keratometry and biometry
The instruments of the eye clinic are applied optics. This lesson explains how the direct and indirect ophthalmoscopes form their images, what the slit lamp, keratometer and applanation tonometer rely on, how the eye is measured before cataract surgery, how the main ophthalmic lasers act on tissue, and what spherical and chromatic aberration are.
The direct ophthalmoscope
In an emmetropic eye the retina lies at the focal point of the eye's own optics. Light leaving a point on the retina therefore emerges from the pupil as parallel rays. An emmetropic examiner looking into the pupil focuses those parallel rays on their own retina and sees the patient's fundus. The direct ophthalmoscope only adds a light shone along the line of sight and a wheel of lenses to cancel any refractive error in patient or examiner.
The patient's eye is acting as a magnifying glass of about +60 D held over its own retina. The magnification of a simple magnifier is its power divided by 4, which compares the view with the same object held at 25 cm.
Magnification of direct ophthalmoscopy = power of the eye / 4For an emmetropic eye of 60 D this is 15 times.Worked example 1. Emmetropic eye: 60 / 4 = 15 times. A myopic eye is more powerful for its length and gives more magnification. A hypermetropic eye gives less, and an aphakic eye, having lost about a third of its power, gives much less. The image is virtual and upright. The price of high magnification is a small field of view, about the size of the disc and its surroundings, and a view through one eye only, so there is no depth.
The indirect ophthalmoscope and slit-lamp lenses
In indirect ophthalmoscopy a strong plus condensing lens is held in front of the patient's eye. It takes the parallel light leaving the eye and focuses it to a real, inverted image hanging in the air between the lens and the examiner. The image is upside down and reversed left to right. The examiner views this aerial image with both eyes through a head-mounted binocular viewer, so the view is stereoscopic.
Magnification of indirect ophthalmoscopy = power of the eye / power of the condensing lensUse 60 D for the eye.Worked example 2. With a 20 D lens, 60 / 20 = 3 times. With a 30 D lens, 60 / 30 = 2 times. With a 28 D lens, 60 / 28 is a little over 2. A stronger lens gives less magnification but a wider field and is easier through a small pupil. The 20 D lens, the standard for adults, shows about 45 degrees of fundus at once, and a 30 D lens about 65 degrees. Exact figures depend on the design.
At the slit lamp the same principle is used with a stronger hand-held lens, most often 90 D or 78 D. The image is again real and inverted, and the slit-lamp microscope supplies further magnification. A 90 D lens on its own gives an image about 0.76 times life size, so with the microscope set at 16 times the fundus is seen at roughly 12 times. The alternative is a contact lens with minus power that cancels the cornea, such as the central part of the Goldmann three-mirror lens, which gives a virtual, upright image.
| Direct ophthalmoscope | Indirect with 20 D | Slit lamp with 90 D | |
|---|---|---|---|
| Image | Virtual, upright | Real, inverted and laterally reversed | Real, inverted and laterally reversed |
| Magnification | About 15 times | About 3 times | Set by the microscope |
| Field of view | Small | Wide | Wide |
| Stereopsis | No | Yes | Yes |
| Best for | Quick look at disc and macula | Peripheral retina, hazy media, children | Detailed disc and macular examination |
The slit lamp
The slit lamp combines a binocular microscope with a bright light that can be narrowed to a thin slit. Light source and microscope pivot around the same point, so the light is always in focus where the examiner is looking. A narrow slit shone obliquely cuts an optical section through the transparent cornea and lens, showing the depth of any opacity. Other lighting methods include diffuse illumination, retroillumination, in which light reflected from the iris or fundus shows up defects from behind, specular reflection for the corneal endothelium, and sclerotic scatter.
Two things cannot be seen with the slit lamp alone. The fundus is out of range because of the power of the cornea and lens, hence the auxiliary lenses above. The drainage angle is hidden by total internal reflection at the cornea, hence the gonioscopy lens.
The keratometer
The keratometer measures the curvature of the front of the cornea. It treats the tear-covered cornea as a convex mirror. An illuminated target of known size, the mire, is reflected in the cornea. A steep cornea gives a small reflected image and a flat cornea a larger one, so the size of the image gives the radius of curvature. Because the eye is never perfectly still, the instrument uses prisms to double the image and the examiner lines up the two copies, which move together. It samples only a small central zone of the cornea, about 3 mm across, in two meridians.
The radius is converted to power using an assumed index of 1.3375, not the true corneal index of 1.376. The lower figure is a fudge that allows for the minus power of the back corneal surface, which the instrument cannot measure.
K (D) = 337.5 / rr is the anterior corneal radius in millimetres.Worked example 3. A radius of 7.5 mm gives 337.5 / 7.5 = 45.00 D. A flatter cornea with a radius of 8.0 mm gives 337.5 / 8 = 42.2 D. Readings of 45.00 D vertically and 43.50 D horizontally mean 1.50 D of with-the-rule corneal astigmatism. Keratometry is used for contact lens fitting, for calculating intraocular lens power and for following keratoconus. Its assumptions fail after corneal laser surgery, when the front surface has been reshaped but the back has not.
The applanation principle
The Goldmann tonometer measures intraocular pressure by flattening a small circle of cornea. The Imbert-Fick law states that the pressure inside an ideal thin-walled, dry sphere equals the force needed to flatten part of its surface divided by the area flattened.
P = F / AP is pressure, F the flattening force, A the area flattened.The cornea is neither infinitely thin nor dry. Its stiffness resists flattening, and the surface tension of the tear film pulls the tonometer tip towards the eye. Goldmann found that when the flattened circle is about 3 mm across these two opposing errors roughly cancel in an average cornea. The exact diameter used is 3.06 mm, an area of about 7.35 square millimetres, which is convenient because a force of 1 gram then corresponds to 10 mmHg. A split-image prism in the tip turns the fluorescein-stained tear ring into two semicircles, and the end point is reached when their inner edges just touch. The assumption about an average cornea is the weak point: a thick cornea gives a falsely high reading and a thin cornea a falsely low one. The glaucoma lessons take this further.
Biometry and intraocular lens power
Removing a cataract removes about a third of the eye's power, and the implant must replace it exactly. Biometry is the set of measurements needed to choose the implant: chiefly the axial length and the corneal power from keratometry.
A-scan ultrasound sends a 10 MHz sound beam along the axis of the eye and times the echoes from the cornea, lens surfaces and retina. Time is converted to distance using the speed of sound, about 1532 m/s in aqueous and vitreous and 1641 m/s in the lens. If the probe touches the cornea it can indent it and give a falsely short length. Immersion scanning through a saline bath avoids this. Optical biometry uses partial coherence interferometry with infrared light (780 nm in the original instrument) to measure to the retinal pigment epithelium. It involves no contact and is the standard method, but the light may not penetrate a very dense or posterior subcapsular cataract, and ultrasound is then needed.
P = A - 2.5L - 0.9KThe SRK formula. P is the implant power for emmetropia (D), A a constant specific to the lens model, L the axial length in mm, K the average keratometry in D.Worked example 4. A = 118, L = 24 mm, K = 44 D. P = 118 - (2.5 x 24) - (0.9 x 44) = 118 - 60 - 39.6 = 18.4 D. Worked example 5. The axial length is wrongly measured as 23 mm. P = 118 - 57.5 - 39.6 = 20.9 D, which is 2.5 D too strong. The patient ends up unexpectedly myopic. An error of 1 mm in length is worth about 2.5 to 3 D of implant power, while 1 D of keratometry error is worth about 1 D, so axial length is the measurement that matters most.
SRK is a regression formula from the 1980s and is no longer used clinically, but it shows the idea: longer eyes and steeper corneas need weaker implants. Modern formulae, such as SRK/T, Hoffer Q, Haigis and Barrett, are based on vergence calculations. Their main task is to predict the effective lens position, where the implant will sit in the eye, which is the one quantity that cannot be measured beforehand.
Lasers
Laser light is of a single wavelength, coherent and highly directional, so it can be focused to a very small spot of high energy. What it does to tissue depends on its wavelength, which decides where it is absorbed, and on how quickly the energy is delivered.
| Laser | Wavelength | Tissue effect | Typical uses |
|---|---|---|---|
| Argon | 488 nm blue and 514 nm green | Photocoagulation: light absorbed by pigment is turned to heat, which burns and scars | Retinal photocoagulation, laser trabeculoplasty |
| Frequency-doubled Nd:YAG | 532 nm green | Photocoagulation | Panretinal and macular laser, including pattern scanning systems |
| Diode | 810 nm infrared | Photocoagulation, penetrating more deeply | Retinopathy of prematurity, cyclodiode treatment of the ciliary body |
| Nd:YAG | 1064 nm infrared | Photodisruption: a very short pulse creates plasma and a shock wave that cuts tissue mechanically | Posterior capsulotomy, peripheral iridotomy |
| Excimer (argon fluoride) | 193 nm ultraviolet | Photoablation: breaks molecular bonds and removes tissue with little heat spread | Reshaping the cornea in PRK and LASIK |
| Femtosecond | About 1053 nm infrared | Ultrashort pulses form plasma that separates tissue with great precision | Cutting the LASIK flap |
Photocoagulation needs a pigment to absorb the light. Melanin in the retinal pigment epithelium and choroid and haemoglobin in blood absorb visible wavelengths well. The yellow xanthophyll pigment of the macula absorbs blue light, so green is preferred to blue for treatment near the fovea. Photodisruption needs no pigment, which is why the infrared Nd:YAG laser can open a clear posterior capsule.
Aberrations
No real lens forms a perfect image. In spherical aberration, rays passing through the edge of a lens with spherical surfaces are bent more than rays through the centre, so they focus at different points and the image is soft. It increases with the size of the aperture, so it is worse with a large pupil and less with a small one. Aspheric lens designs, whose curvature changes towards the edge, reduce it.
In chromatic aberration, short wavelengths are refracted more than long ones, so blue light focuses in front of red. In the eye this is the basis of the duochrome test. In instruments it is reduced by an achromatic doublet, a pair of lenses made from glasses of different dispersion whose colour errors cancel. A smaller aperture also reduces the blur from both aberrations, up to the point where diffraction takes over.
Summary
- The direct ophthalmoscope gives a virtual, upright image magnified about 15 times with a small field.
- The indirect ophthalmoscope gives a real, inverted, laterally reversed image, with magnification of 60 divided by the condensing lens power.
- A stronger condensing lens gives less magnification and a wider field.
- The keratometer treats the central cornea as a convex mirror and converts radius to power using an index of 1.3375: K = 337.5 / r in mm.
- Applanation tonometry rests on the Imbert-Fick law, P = F / A, with a flattened diameter of 3.06 mm.
- A-scan ultrasound and optical biometry measure axial length, and corneal contact in A-scan can give a falsely short reading.
- The SRK formula is P = A - 2.5L - 0.9K, and 1 mm of axial length error is worth about 2.5 to 3 D of implant power.
- Argon and frequency-doubled Nd:YAG lasers photocoagulate, the 1064 nm Nd:YAG laser photodisrupts and the 193 nm excimer laser photoablates.
- Spherical aberration arises because peripheral rays are refracted more than central rays and is reduced by a small pupil.
- Chromatic aberration arises because blue light is refracted more than red.
Check yourself