Slit lamp examination, tonometry and gonioscopy

The slit lamp is the ophthalmologist's main instrument: a microscope joined to a light source whose beam can be shaped, angled and filtered. This lesson explains what each way of lighting the eye shows, how fluorescein is used, how intraocular pressure is measured and why the reading can mislead, and how the drainage angle is viewed and graded.

10 minute read2 linked questions

The instrument

A slit lamp has three parts: a binocular microscope, an illumination arm that projects a bright beam, and a frame with a chin rest that holds the patient still. The microscope gives a stereoscopic view at magnifications from about 6 to 40 times. The illumination arm swings independently of the microscope, and controls on it change the height and width of the beam and bring filters into the light path. Both arms are focused on the same point, so whatever is in focus is also lit.

The examination follows the anatomy from front to back: lids and lashes, conjunctiva and sclera, cornea, anterior chamber, iris, lens and anterior vitreous. The cornea and lens together are too powerful to let the microscope focus further back than the front of the vitreous, so the fundus and the drainage angle each need an extra lens.

Ways of lighting the eye

The value of the slit lamp lies in the fact that transparent tissues scatter a little light. A narrow beam passing obliquely through the cornea lights up a thin slice of it, like a sunbeam crossing a dusty room. Different arrangements of the beam pick out different abnormalities.

TechniqueHow it is set upWhat it shows
DiffuseWide beam through a diffuser, low magnificationGeneral survey of lids, conjunctiva and ocular surface
Direct focal: optical sectionVery narrow beam at an angle, focused where the microscope is focusedA cross-section of the cornea or lens, so that the depth of an opacity or foreign body can be judged
Direct focal: small bright beamShort, narrow, bright beam in a dark room, aimed through the pupilCells and flare in the aqueous
RetroilluminationLight bounced back from the iris or the red reflex of the fundusCorneal oedema and deposits, lens opacities, and iris transillumination defects
Sclerotic scatterBeam aimed at the limbus, so light travels within the cornea by internal reflectionSubtle corneal opacities, which interrupt the light and glow
Specular reflectionAngle of the microscope set equal to the angle of the lightThe corneal endothelial cell mosaic
Slit lamp illumination techniques

In an optical section of the cornea the surface nearest the light is the epithelium and the far surface is the endothelium. Keratic precipitates, the clumps of inflammatory cells seen in uveitis, sit on the far surface.

Fluorescein, the Seidel test, cells and flare

Fluorescein is an orange dye that absorbs blue light and emits green. With the cobalt blue filter in the beam, anything holding fluorescein glows green. Intact corneal epithelium repels the dye. Where epithelium is missing, the dye pools and enters the stroma, so abrasions, ulcers and the branching dendritic ulcer of herpes simplex keratitis stain bright green. Fluorescein also shows up the tear film for tonometry and for measuring tear break-up time.

The Seidel test looks for a leak of aqueous through a full-thickness wound of the cornea or sclera. Concentrated fluorescein is painted over the suspect area. Concentrated dye looks dark orange and fluoresces poorly. If aqueous is leaking, it dilutes the dye, and a stream of contrasting colour runs down from the wound under blue light. It is usually described as bright green against the dark concentrated dye, although some sources describe a dark rivulet through a green film. A positive test means the globe is open: after trauma, a perforated ulcer, or a leaking surgical wound or drainage bleb. It should not be done when the globe is obviously ruptured.

Normal aqueous is optically empty. In anterior uveitis the blood-aqueous barrier breaks down. Cells are individual white cells seen drifting in the beam like dust motes. Flare is haze of the beam itself, caused by leaked protein scattering light. Both are sought with a small, bright beam in a darkened room and graded on standard scales, which the uveitis chapter covers. Cells indicate active inflammation. Flare can persist after inflammation has settled.

Estimating the angle: the van Herick test

The van Herick test estimates how deep the anterior chamber is at its edge, without touching the eye. A narrow beam is directed at the peripheral cornea just inside the limbus, at about 60 degrees to the microscope. The examiner compares the dark gap between the back of the cornea and the iris with the thickness of the corneal section.

GradePeripheral chamber depth compared with corneal thicknessRisk of angle closure
4Equal to or greater than corneal thicknessVery unlikely
3A quarter to a halfUnlikely
2About a quarterPossible
1Less than a quarterLikely
0No gapClosed
Van Herick grades

The test is a screen. It shows no angle structures and cannot replace gonioscopy. NICE NG81 asks for gonioscopy when glaucoma is diagnosed, and accepts the van Herick test where gonioscopy is not available or not possible.

Goldmann applanation tonometry

Intraocular pressure cannot be measured directly without entering the eye, so it is inferred from how hard the cornea is to deform. Goldmann applanation tonometry is the reference method and the one NICE specifies for diagnosing and monitoring glaucoma. It rests on the Imbert-Fick principle: for an ideal thin, dry, perfectly flexible sphere, the pressure inside equals the force needed to flatten part of its surface divided by the area flattened.

P = F / AP is pressure, F the flattening force and A the area flattened.

The cornea is not ideal. It is stiff, which resists flattening and adds to the force needed. It is wet, and surface tension in the tear film pulls the tonometer tip onto the eye, reducing the force needed. Goldmann found that when the flattened circle is 3.06 mm in diameter these two effects roughly cancel. At that diameter there is also a convenient conversion: the force in grams multiplied by 10 gives the pressure in mmHg. Worked example: if 1.6 g flattens the 3.06 mm circle, the pressure is 16 mmHg.

In practice the cornea is anaesthetised and fluorescein instilled. A prism in the tonometer tip splits the view of the fluorescein ring around the flattened area into two green semicircles, one above the other. The dial is turned until the inner edges of the two semicircles just touch, and the pressure is read from the dial. The semicircles pulse with the heartbeat, and the endpoint is taken at the middle of the pulsation.

Reading falsely highReading falsely low
Thick central corneaThin central cornea, including after laser refractive surgery
Too much fluorescein (wide semicircles)Too little fluorescein (thin semicircles)
Corneal scarringCorneal oedema
Pressure on the globe from the examiner's fingers or squeezing lidsMeasuring over a contact lens
Straining (a Valsalva effort) or a tight collarRepeated measurements in quick succession
Sources of error in Goldmann tonometry

Corneal thickness matters most. Goldmann calibrated the instrument for a cornea about 520 micrometres thick. A thicker cornea needs more force to flatten and reads high. A thinner one reads low, so a thin cornea can hide a dangerous pressure. This is why NICE requires central corneal thickness to be measured when glaucoma or ocular hypertension is diagnosed. With marked corneal astigmatism the semicircles become elliptical and the prism must be rotated to a marked position to compensate.

Other tonometers

InstrumentPrincipleWhere it is used
PerkinsHand-held Goldmann-type applanationPatients who cannot sit at the slit lamp, children, examination under anaesthesia
Non-contact (air puff)A rising jet of air flattens the cornea; the moment of flattening is detected opticallyOptometric screening; no anaesthetic or contact needed
Rebound (iCare)A tiny probe bounces off the cornea; its deceleration gives the pressureChildren and home monitoring; no anaesthetic needed
Tono-PenHand-held electronic device with a small tip and strain gaugeScarred or irregular corneas, bedside use
SchiotzIndentation: a weighted plunger sinks further into a softer eyeLargely historical; still used where resources are limited
Dynamic contour (Pascal)Tip shaped to the cornea with a pressure sensorDesigned to depend less on corneal properties
Tonometers other than slit-lamp Goldmann

The non-contact and rebound instruments are also affected by corneal thickness and stiffness. Air-puff readings tend to overestimate low pressures and underestimate high ones, and several readings are averaged because pressure varies with the pulse. A high screening reading is therefore confirmed with Goldmann tonometry before a diagnosis is made. Mean pressure in European populations is about 15.5 mmHg, with 21 mmHg roughly two standard deviations above the mean, and it varies by a few mmHg through the day.

Gonioscopy

The drainage angle lies where the peripheral iris meets the cornea. It cannot be seen by looking straight into the eye, because light leaving the angle strikes the front surface of the cornea so obliquely that it is totally internally reflected back into the eye. A gonioscopy lens placed on the cornea, with fluid between, abolishes the cornea-air interface and lets the light out.

Indirect lenses contain mirrors and are used at the slit lamp. Each mirror shows the angle opposite to it. The Goldmann lens is large and needs a coupling gel. The four-mirror Zeiss, Posner and Sussman lenses have a small contact area, so the examiner can press on the central cornea and push aqueous into the angle. This is indentation gonioscopy. An angle that is merely closed by contact (appositional closure) opens with indentation. One sealed by scar adhesions between iris and meshwork, called peripheral anterior synechiae, does not. Direct lenses such as the Koeppe give an upright view with the patient lying down and are used mainly in theatre.

In an open angle four landmarks are seen, from the cornea back towards the iris root.

  1. Schwalbe line: a fine pale line marking the end of Descemet membrane.
  2. Trabecular meshwork: a band whose posterior part, lying over Schlemm canal, is pigmented to a variable degree. Schlemm canal itself is seen only when blood refluxes into it.
  3. Scleral spur: a white ridge of sclera to which the ciliary muscle attaches.
  4. Ciliary body band: a grey or brown band, visible only in a wide angle.
3D atlasThe drainage angle: the structures a gonioscopy lens brings into view.

The more of these structures that can be seen, the wider the angle. The Shaffer system grades the angle by its estimated width: grade 4 is 35 to 45 degrees and wide open, grade 3 is 20 to 35 degrees, grade 2 is about 20 degrees with closure possible, grade 1 is 10 degrees or less with closure probable, and grade 0 is closed. Gonioscopy also shows new vessels in the angle, angle recession after blunt trauma, and the heavy pigment of pigment dispersion and pseudoexfoliation.

Summary

  • The slit lamp is a stereoscopic microscope with an independently angled beam; an optical section shows the depth of a lesion in the cornea or lens.
  • Retroillumination shows lens opacities and iris transillumination, sclerotic scatter shows faint corneal opacities, and specular reflection shows the endothelium.
  • Fluorescein stains epithelial defects green under the cobalt blue filter.
  • A positive Seidel test, a stream of aqueous diluting the dye, means aqueous is leaking through a full-thickness wound.
  • Cells are inflammatory cells in the aqueous and flare is protein haze; both mean breakdown of the blood-aqueous barrier.
  • Van Herick compares peripheral chamber depth with corneal thickness; a quarter or less suggests a narrow angle.
  • Goldmann tonometry uses the Imbert-Fick principle, flattening a 3.06 mm circle, with force in grams times 10 giving mmHg.
  • Thick corneas give falsely high Goldmann readings and thin corneas falsely low ones.
  • The angle is hidden by total internal reflection and is seen only through a gonioscopy lens.
  • An open angle shows Schwalbe line, trabecular meshwork, scleral spur and ciliary body band; indentation separates appositional closure from synechiae.

Check yourself

Quick recall on this lesson. These are easier than the question bank.

1. What diameter of cornea is flattened in Goldmann applanation tonometry?
2. How does a thin central cornea affect a Goldmann reading?
3. Why can the drainage angle not be seen without a gonioscopy lens?
4. Which gonioscopic landmark lies most anteriorly?